SP-UV1100 Spectrophotometer

SP-UV1100 Spectrophotometer

Brand: DLAB
Product Code: SP-UV1100
Availability: In Stock

• User Friendly Information Rich LCD Display. The large 128 x 64 dots, backlit LCD screen with adjustable brightness control displays a large array of data also in graphical format.

• On-Board Data Storage. The System can save the test results, up to 200 groups of data and 200 standard curves in the RAM memory.

• Data can be restored after a sudden power failure.

• Auto setting wavelength. To set wavelength automatically, to calibrate the system through arrow keys, to avoid operation errors.

• Tungsten & Deuterium lamps switching is automatic. Selectable from a wavelength range of 190 to 1100nm.

• PC control through application software. The spectrophotometer is fully equipped and capable of executing all functions in stand-alone mode. The optional application software Wave Professional provides computer control for the spectrophotometer ( through the builtin USB port), apart from other functions such as spectrum scan, Time scan ( kinetics), multiwavelength scanning& Nucleic Acid /Protein measurement.(SP-V1100/SP-UV1100)

• Large sample compartment. To accommodate 5 – 100mm path length cuvettes with optional holders. A variety of optional accessories are available.

 

 

Model

SP-V1000

SP-UV1000

SP-V1100

SP-UV1100

Optical System

Single Beam

Light Source

Tungsten lamp

Tungsten lamp&

Deuterium lamp

Tungsten lamp

Tungsten lamp &Deuterium lamp

Spectral Bandwidth

4.0nm

2.0nm

Wavelength Range

325~1000nm

200~1000nm

320~1100nm

190~1100nm

Wavelength Accuracy

±2nm

±0.5nm

Wavelength

Repeatability

≤1nm

≤0.3nm

Photometric Range

-0.3 ~ 3.0 A,0 ~ 200 %T

Photometric Accuracy

±0.003 A @ 0.5 A,±0.006 A @ 1.0 A,±0.5 %T @ 100

±0.002 A @ 0.5 A,±0.004 A @ 1.0A,±0.3 %T @ 100 %T

Photometric

Repeatability

≤0.0015 A @ 0.5 A,≤0.003 A @ 1.0 A,≤0.25

≤0.001 A @ 0.5 A,≤0.002 A @ 1.0A,≤0.15 %T @ 100 %T

Stability

≤0.002 A/h @ 500 nm ( after preheat 2 hours)

Stray Light

≤0.2@ 360 nm

≤0.2@ 220& 360 nm

≤0.05@ 360nm

≤0.05@ 220 & 360 nm

Sample Compartment

4-position, 10mm path length cuvette

Display

LCD

Output

USB Port & Parallel Port ( Printer)

Dimensions( L×W×H)

490×360×210 mm

Weight

10kg

12kg

10kg

12kg